Cookies on the OMEGA websites We use cookies on this website, these cookies are essential for the website to work correctly.If you continue without changing your settings, we'll assume that you are happy to receive all cookies on this website.To find out more information about these cookies please click here.
The DOTX441 dissolved oxygen analyzer is
supplied in a compact cast aluminum case. SAE 323,
with low oxidation grade, anticorrosion treatment and
finished with electrostatic epoxy paint. The DOTX441 is
compact, light weight, rugged and IP-68 rated. The units include
local display, transmitter output and controller function all
programmed via a 3-button side panel with an easy self guided
menu. The DOTX441 can be mounted on a 51 mm (2") tube or flat
surface. Electrical connections are on the terminal bar located in
the back housing of the instrument with cable entrance through
two cable compression fittings of 13 mm (½") BSP. Common
applications include food and drink, pulp and paper, chemical,
sanitation, water and wastewater treatments.
SPECIFICATIONS Case Cast: Aluminum SAE 323 Finishing: Electrostatic epoxy paint Electrical Connection Connector Bar: Conexell type Cable Inlet Cable Knockout: 13 mm (½") (2 x) Assembly: 51 mm (2") tube or flat surface Identification: SS tag Enclosure: NEMA 7X (IP68) Power Consumption: 3.5 VA Weight: 1.9 kg (4.3 lb) Electrical Power: 90 to 240 Vac (50/60 Hz) Working Temperature: 5 to 40°C (41 to 104°F) Relative Humidity: 20 to 80%
Analyzer/Transmitter Display Alphanumeric: 2 lines x 16 characters Range: 0 to 60 mg O2/L Resolution: 0.1 or 0.01 Relative Precision: 0.02% (full scale) Working Temperature: 50°C (122°F) Auto Temperature Compensation: 50°C (122°F) Measuring Principle: Clark polarographic cell Response Time: 20 s (programmable) Barometric Pressure Compensation: Up to 10.000 m Signal Output Analogic: 4 to 20 mA with programmable output range for all reading range Galvanic Isolation: 2000 Vac (by opto coupler) Line Resistance: 1 kΩ
Controller Actuation Type Frequency: Modulation, P+D Set-Points: 2 independent from 0 to 100% of scale Outputs: 2 control ON/OFF, programmable for alarm or actuation P+D Control for Automatic Cleaning: Timer ON/OFF, for periods up to 99 seconds, and in intervals up to 99 hours
Cell Type: Immersion Body material: polypropylene (PP) or 316 SS Maximum Temperature: (@ 1 ATM) 50°C (122°F) Body Length: 1.5 m (59"), standard Auto Cleaning System: Activated by a solenoid valve
† All amounts shown in EUR, GBP, or USD Note: Comes complete with 304 SS clamps with nuts and bolts for mounting, and operator’s manual. Sensor sold separately. Ordering Example: (1) DOTX441 Dissolved oxygen analyzer, transmitter and controller, £879.06