Description
The DP63100 series analogue process panel meters offer many features and performance capabilities to suit a wide range of industrial applications. The DP63100 has a universal input to handle various input signals including DC voltage/current, process, resistance and temperature. The DP63100-S model is for strain Gauge/Load-Cell inputs.
The optional plug-in output cards allow the opportunity to configure the meters for present applications, while providing easy upgrades for future needs. The meters have a dual line display with a large 18 mm high, tri-colour, 6 digit top display line and a 8.9 mm high, 9 digit green bottom display line. The meter provides a MAX and MIN reading memory with programmable capture time. The capture time is used to prevent detection of false max or min readings which may occur during start-up or unusual process events.The signal totaliser (integrator) can be used to compute a time-input product. This can be used to provide a readout of totalised flow or calculate service intervals of motors, pumps, etc.
The meters have up to four setpoint
outputs, implemented on plug-in
option cards. The plug-in cards
provide dual FORM-C relays, quad
FORM-A, or either quad sinking or
quad sourcing open collector logic
outputs. The setpoint alarms can be
configured to suit a variety of control
and alarm requirements.
Communication and bus capabilities
are also available as option cards.
These include RS232 and RS485.
The meter's can be programmed
to use MODBUS® protocol. With
MODBUS, the user has access to
most configuration parameters.
Readout values and setpoint alarm
values can be controlled through
the bus. Additionally, the meter has
a feature that allows a remote
computer to directly control the
outputs of the meter. A built-in USB
programming port. With a Windows®
based program, configuration data
can be downloaded to the unit
without the need of any additional
option cards.
A linear analogue DC output
signal is available as an optional
plug-in card. The card provides
either 20 mA or 10V signals. The
output can be scaled independent
of the input range and can track
either the input, totalizer, max or
min readings. The meter has been
specifically designed for harsh
industrial environments. With
an IP65 sealed front lens and
extensive testing of noise effects
and CE requirements, the meter
provides a tough reliable
application solution.
SPECIFICATIONS
Display: Negative image LCD
Top Line: 6 digit, 18 mm, with tri-colour backlight (red, green or orange)
Display Range: -199999 to 999999;
Bottom Line: 9 digit, 8.9 mm, with green backlight
Display Range: -199,999,999 to 999,999,999
Power
AC Power: 50 to 250 Vac, 50/60 Hz, 14 VA
DC Power: 21.6 to 250 Vdc, 8 W
Isolation: 2300 Vrms for 1 min to all inputs and outputs. KEYPAD: 3 programmable function keys, 5 keys total
A/D Converter: 24 bit resolution
Update Rates: A/D conversion rate; programmable 5 to 160 readings/sec
Thermocouple Inputs
Readout:
Scale: °F or °C
Offset Range: -199,999 to 999,999 display units
Input Impedance: 20 MO
Lead Resistance Effect: 0.03 µV/Ω
Max Continuous Overvoltage: 30V
RTD Inputs
Type: 3 or 4 wire, 2 wire can be compensated for lead wire resistance
Excitation Current: 100 Ω
Range: 136.5 µA ±10%
10 Ω Range: 2.05 mA ±10%
Lead Resistance: 100 Ω
Range: 10 Ω/lead max
10 Ω Range: 3 Ω/lead max
Max Continuous Overload: 30V
Excitation Power: Jumper selectable
Transmitter Power: ±18 Vdc @ 50 mA
Reference Voltage: ±2 Vdc, ±2%
User Inputs
Two Programmable User Inputs Max Continuous Input: 30 Vdc
Isolation to Sensor Input Common: Not isolated
Response Time: 12 msec max
Totaliser
Time Base: Second, minute, hour, or day
Batch: Can accumulate (gate) input display from a user input
Time Accuracy: 0.01% typical
Decimal Point: 0 to 0.0000
Scale Factor: 0.001 to 65.000
Low Signal Cut-Out: -199,999 to 999,999
Total: 6 digits on line 1; 9 digits on line 2
Custom Linearization
Data Point Pairs: Selectable from 2 to 16
Display Range: -199,999 to 999,999
Decimal Point: 0 to 0.0000
Memory: Nonvolatile FRAM memory retains all programmable parameters and display values
Environmental Conditions
Operating Temperature Range: 0 to 50°C or 0 to 45°C with all three plug-in cards installed
Storage Temperature Range: -40 to 60°C
Operating and Storage Humidity: 0 to 85% max RH non-condensing
Altitude: Up to 2000 m
Connections: High compression age-clamp terminal block
Construction: This unit is rated for NEMA 4X (IP65) outdoor use IP20 touch safe. Installation Category II, pollution degree 2. One piece bezel/case. Flame resistant. Synthetic rubber keypad. Panel gasket and mounting clip included.
Dimensions
Case: 44.45 H x 91.44 W x 105 mm D
Bezel: 49.53 x 96.52 mm
Panel Cutout: 1/8 DIN, 45 x 92 mm
Weight: 226.8 g