pH meter $amp; transmitter | PHTX-100

pH meter $amp; transmitter

£522.91 PHTX-100

CE
  • Loop Powered, 4 to 20 mA
  • Non-Volatile Memory for settings and Calibration Data
  • Output Current Indicator for Out-of-Range and Hold Conditions
  • Large Display Indicates ph or mV and Temperature
  • Programmable Probe Clean Reminder (Programmable from 1 to 250 Days)
Industrial pH Instrumentation and Electrodes - View related products

Description

The PHTX-100 pH/mV transmitter is a loop-powered 4 to 20 mA transmitter with a wide range of useful features and programmable options. Five buffer choices are available in two-point calibration mode. The 4 to 20 mA output is saleable in pH or mV units and has 500V input to output isolation.

Temperature is always displayed and is selectable in °C or °F. All stored parameters and calibration data are retained in the non-volatile memory during transmitter power down or power loss. A useful probe clean reminder icon flashes when it is time to clean the probe. Probe cleaning interval is user programmable from 1 to 250 days.

The PHTX-100 can be used for wall mounting, panel mounting, pipe mounting and DIN-rail mounting. Cable glands for inlet/outlet to box are included as well as all mounting hardware. The case meets IP65 (NEMA 4X) ratings for ingress protection.

Specifications
Measuring Range (pH): 0.00 to 14.00 pH, 0.01 pH resolution, ±0.01 accuracy
Measuring Range: (ORP/mV) -1999 to 1999 mV, 1 mV resolution, ±2 mV accuracy
Temperature Range: -20 to 110°C (-4 to 230°F) 0.1°C/°F resolution
Current Output Range: 2.00 to 24 mA, 0.01 mA resolution, ±0.005 mA accuracy, 500V input/output isolation
Enclosure: NEMA 4X (IP65), ABS case with push-button control pad
Dimensions: 99 H x 99 W x 75 mm D
Weight: 0.45 kg approx.
Mounting Options: Wall, panel, DIN rail and pipe
Cable Entries: 2, cable glands included
Ambient Temperature: Transmitter service, 0 to 60°C
Ambient Humidity: 0 to 95%, non-condensing
Location: Designed for non-hazardous areas
Temperature Input: 2-wire Pt1000 RTD with automatic compensation
Maximum Sensor-to-Transmitter Distance: 9 metres
Power Requirement: 12 to 24 Vdc, maximum 600Ω load
Place Order(Specify Model Number)
Part Number/ Desc.
Qty
Description
PHTX-100
£522.91
Availability: 2 weeks
 
2-wire, 4 to 20 mA pH transmitter with display
All amounts shown in EUR, GBP, or USD
Note: Comes complete with mounting hardware and operator's manual.