What does the pH measure?
The pH value of a substance is directly related to the ratio of the hydrogen ion [H+] and the hydroxyl ion [OH-] concentrations.
The quantitative information provided by the pH tester expresses the degree of the activity of an acid or base in terms of hydrogen ion activity.
- If the H+ concentration
is greater than OH-, the material is acidic; i.e., the pH measurement is less than 7.
- If the OH- concentration is greater than H+, the material is basic,
with a pH value greater than 7.
- If equal amounts of H+ and OH- ions are present, the material is neutral, with a pH of 7.
Acids and bases have free hydrogen and hydroxyl ions, respectively. The relationship between hydrogen ions and hydroxyl ions in a given solution is constant for a given set of conditions, either one can be
determined by knowing the other.
How to measure pH?
A rough indication of pH can be obtained using pH papers or indicators, which change color as the pH level varies.
These indicators have limitations on their accuracy, and can be difficult to interpret correctly in colored or murky samples.
More accurate pH measurements are obtained with a digital pH meter. A pH tester system consists of three parts: a pH probe,
a reference pH electrode, and a high input impedance meter.
The pH electrode can be thought of as a battery, with a voltage that varies with
the pH of the measured solution. The pH probe is a hydrogen ion sensitive glass bulb, with a millivolt output that varies
with the changes in the relative hydrogen ion concentration inside and outside of the bulb.
The reference electrode output does not vary
with the activity of the hydrogen ion.
The pH electrode has very high internal resistance, making the voltage change with pH difficult to
measure. The input impedance of the pH meter and leakage resistances are therefore important factors.
The pH meter is basically a high
impedance amplifier that accurately measures the minute electrode voltages and displays the results directly in pH units on either an analog
or digital display.
In some cases, voltages can also be read for special applications or for use with ion-selective or Oxidation-Reduction
Potential (ORP) electrodes.
pH Electrodes
pH
electrodes are constructed from a special composition glass which
senses the hydrogen ion concentration. This glass is typically
composed of alkali metal ions. The alkali metal ions of the glass
and the hydrogen ions in solution undergo an ion exchange reaction,
generating a potential difference.
In a combination pH probe,
the most widely used variety, there are actually two electrodes
in one body. One portion is called the measuring pH electrode, the
other the reference electrode. The potential generated at the
junction site of the measuring portion is due to the free hydrogen
ions present in solution.
The potential of the reference portion is produced by the internal
element in contact with the reference fill solution. This potential
is always constant. In summary, the measuring pH electrode delivers
a varying voltage and the reference electrode delivers a constant
voltage to the meter.
The voltage signal produced by the pH probe
is a very small, high impedance signal. The input impedance requires
that it be interfaced only with equipment with high impedance
circuits.
pH electrodes are available in a variety of styles for both laboratory
and industrial applications. All are composed of glass and are
therefore subject to breakage.
Electrodes are designed to measure
mostly aqueous media. They are not designed to be used in solvents,
such as CCI4, which does not have any free hydrogen ions.
The
pH electrode, due to the nature of its construction, needs to
be kept moist at all times. In order to operate properly, glass
needs to be hydrated.
Hydration is required for the ion exchange
process to occur. If an pH sensor should become dry, it is best
to place it in some tap water for a half hour to condition the
glass.
pH
electrodes have junctions which allow the internal fill solution
of the measuring electrode to leak out into the solution being
measured. This junction can become clogged by particulates in
the solution and can also facilitate poisoning by metal ions present
in the solution.
If a clogged junction is suspected it is best
to soak the sensor in some warm tap water to dissolve the material
and clear the junction. pH testers should always be stored
in a moistened condition. When not in use it is best to store
the electrode in either buffer 4.0 or buffer 7.0.
Even if an electrode is not used it still ages.
On the shelf, the pH probe should last approximately a year if
kept in a moistened condition.
Electrode demise can usually be
characterized by a sluggish response, erratic readings or a reading
which will not change. When this occurs an electrode can no longer
be calibrated. pH electrodes are fragile and have a limited lifespan.
How long an electrode will last is determined by how well the
probe is maintained and the pH application.
The harsher the system,
the shorter the lifespan. For this reason it is always a good
idea to have a back-up pH sensor on hand to avoid any system down
time.
Calibration is also an important part of electrode maintenance.
This assures not only that the electrode is behaving properly
but that the system is operating correctly.
pH meter calibration
pH
electrodes are like batteries; they run down with time and use.
As an electrode ages, its glass changes resistance. This resistance
change alters the electrode potential. For this reason, electrodes
need to be calibrated on a regular basis.
Calibration in pH buffer
solution corrects for this change. Calibration of any pH equipment
should always begin with buffer 7.0 as this is the "zero point."
The pH scale has an equivalent mV scale. The mV scale ranges from
+420 to -420 mV. At a pH of 7.0 the mV value is 0. Each pH change
corresponds to a change of ±60 mV. As pH values become more acidic
the mV values become greater.
For example, a pH of 4.0 corresponds
to a value of 180 mV. As pH values become more basic the mV values
become more negative; pH=9 corresponds to -120 mV. Dual pH calibration
using buffers 4.0 or 10.0 provides greater system accuracy.
Here is a general method for most digital pH meters. Some pH testers require slightly different techniques. Please read the instructions for
their particular procedures.
- The temperature setting on the meter must correspond to the temperature of the buffers used, or an automatic temperature
compensator must be employed.
- Turn pH meter to “pH” or “ATC” if automatic temperature compensation is used.
- Place clean electrode into fresh, room temperature pH 7.00 buffer.
- Adjust the pH reading to exactly 7.00 using the ZERO OFFSET,STANDARDIZED or SET knob.
- Rinse the electrode with distilled or deionized water. (This would be the procedure for a one-point calibration. Continue
through step 8 for a two-point calibration.)
- Place electrode into the second buffer, either pH 4.00 or pH 10.00.
- Adjust the pH reading to display the correct value using the SLOPE, CALIBRATE, or GAIN controls (coarse adjust).
- Adjust the pH reading to read the correct value using the SLOPE knob (fine adjust).
Buffer Solutions
Buffers are solutions that have constant pH values and the ability to resist changes in that pH level. They are used to calibrate pH
measurement systems (electrode and meter). There can be small differences between the output of one electrode and another, as well as changes
in the output of electrodes over time. Therefore, the system must be periodically calibrated. Buffers are available with a wide range of pH values,
and they come in both premixed liquid form or as convenient dry powder capsules. Most pH testers require calibration at several specific pH values.
One calibration is usually performed near the isopotential point (the signal produced by an electrode at pH 7 is 0 mV at 25°C), and a second is
typically performed at either pH 4 or pH 10. It is best to select a buffer as close as possible to the actual pH value of the sample to be measured.
Choose the right pH meter
Portable pH Meters
These pH meters offer full functionality in a portable size for use in the lab or field. Features available include RS232
output, Data Hold, °C or °F selectable, auto shutoff, overload indication and automatic or manual temperature compensation, and more.
Benchtop pH Meters
Benchtop pH meters are usually ideal for laboratory, industrial, and manufacturing applications. Models offer mV, ion and temperature measurement and
range from economical to high-end for low-tolerance measurement needs.
pH Electrodes
pH electrodes are available in a variety of styles for both laboratory and industrial applications. No matter their status, they
are all composed of glass and are therefore subject to breakage. Electrodes are designed to measure mostly aqueous media.
They are not designed to be used in solvents, such as CCI4, which does not have any free hydrogen ions.
pH panel Controller
Its small size allows for easy installation and will accommodate any pH electrode with BNC connector. Temperature compensation is either manual or automatic with use of any 1000Ω platinum RTD.
Frequently Asked Questions
Effects of Temperature on pH
Temperatures Above 25°C: temperature compensation lowers high pH and raises low pH, resulting in value closer to neutral.
Temperatures Below 25°C: temperature compensation raises high pH (more basic) and lowers low pH (more acidic), resulting
in values further away from neutral.
Whether or not temperature compensation needs be used is a matter of the needed
pH accuracy. For example, if the accuracy requirement is ±0.1 pH, at a pH of 6 and a temperature of 45°C (113°F), the error
is 0.06, well within the accuracy requirements. On the other hand, with the same ±0.1 pH accuracy requirement, operating at pH 10 and 55°C (131°F)
would give an error of 0.27 pH and compensation should be used.
When compensation is required, it can be done in one of two ways. If the temperature fluctuates, then an automatic compensator should be used. If the
temperature is constant within several degrees C, then a manual compensator can be used. If no compensator is needed, a fixed resistor can be
installed across the temperature compensator terminals.
Any of the above devices–automatic compensation, manual compensation or fixed resistor–operate as a function of the pH meter’s electronic circuit. As such,
information and parts should be obtained from the meter manufacturer. If automatic compensators are used, they should always be at the same location
as the pH electrode.
When electrodes are calibrated in buffer, the temperature compensator also should be in the buffer. In a similar way, a manual
temperature compensator should be adjusted to reflect the temperature to which the pH electrode is exposed during both calibration and operation.
Temperature Compensation
Temperature compensation is contained within the instrument, because pH electrodes and measurements are temperature sensitive.
The temperature compensation may be either manual or automatic. With manual compensation, a separate temperature measurement is required,
and the pH meter manual compensation control can be set with the approximate temperature value. With automatic temperature compensation (ATC),
the signal from a separate temperature probe is fed into the pH sensor, so that it can accurately determine pH value of the sample at that temperature.
Temperature compensation is contained within the instrument, because pH electrodes and measurements are temperature sensitive.
The temperature compensation may be either manual or automatic. With manual compensation, a separate temperature measurement is required,
and the pH meter manual compensation control can be set with the approximate temperature value. With automatic temperature compensation (ATC),
the signal from a separate temperature probe is fed into the pH meter, so that it can accurately determine pH value of the sample at that temperature.
Challenges of pH Measurement Applications
No pH complications are created equal. The list below illustrates the types of problems that you can expect when measuring pH and how to handle them.
- Instrumentation is frequently the source of disturbance for pH systems, through repeatability error, measurement noise, or valve hysteresis.
- In-line pH loops will oscillate, regardless of controller modes and tuning, if setpoints are on the steep parts of the titration curves.
- pH electrode submersion assemblies with unencapsulated terminations below the liquid surface will eventually have wet terminations.
- Reagent control valves that are not close-coupled to the injection point on in-line systems will cause reagent delivery delays large
enough to describe the tools of your trade n words that may seem foreign.
- You need either a flow meter or a seer to diagnose reagent delivery problems.
- Flow feedforward signals should be multiplied by pH controller outputs and employed to operate reagent valves directly or to establish reagent
flow control setpoints.
- Transportation delays to pH electrodes in analyzer houses will exceed mixing deadlines - such that increasing comfort in checking the
electrodes is offset by decreasing comfort in checking trend recordings.
- Injection electrodes should be preferred to sample holder assemblies whenever possible to reduce maintenance problems and improve
response times - but not all injection electrodes are created equal.
- Large tanks are fine if you don't have to control them; use the volume upstream to reduce reagent consumption or downstream to reduce
control error. If you can't make-up your mind where to use one, put it downstream.
- Install one or three but never two electrodes for a pH measurement.
Rangeability
and Sensitivity
One
basic source of difficulty is that - as countless articles, technical
papers, and textbooks point out - the pH scale corresponds to
hydrogen ion concentrations from 100 to 10-14 moles per liter.
No other common measurement covers such a tremendous range.
Another
intrinsic constraint is that pH sensors can respond
to changes as small as 0.001 pH, so instruments can track hydrogen
ion concentration changes as small as 5x10-10 moles per liter
at 7 pH. No other common measurement has such tremendous sensitivity.
Rangeability and sensitivity limitations can be overcome by approaching
the setpoint in stages, using successively smaller control valves
with high performance positioners.