ANSI Code |
Alloy Combination |
|
Maximum Useful Temperature Range ++ |
Maximum Thermocouple Grade Temperature Range |
EMF (mV) Over Max. Temperature Range |
Standard Limits of Error** (above O°C) |
Special Limits of Error** (above O°C) |
International IEC 584-3 |
Comments Environment — Bare Wire |
IEC Code |
|||
+ Lead | - Lead | Thermocouple Grade |
Extension Grade |
||||||||||
K | CHROMEGA® NICKELCHROMIUM Ni-Cr |
ALOMEGA® NICKEL-ALUMINUM Ni-AI (magnetic) |
Thermocouple Grade: –328 to 2282°F –200 to 1250°C Extension Grade: 32 to 392°F 0 to 200°C |
–454 to 2501°F –270 to 1372°C |
–6.458 to 54.886 |
greater of 2.2°C or 0.75% |
greater of 1.1°C or 0.4% |
Clean Oxidizing and Inert. Limited Use in Vacuum or Reducing. Wide Temperature Range, Most Popular Calibration |
K |
++ Except as further restricted by temperature limits for T/C diameter and insulation tables that follow.
* Not official symbol or standard designation.
** For controlled initial accuracy below 0°C, ask OMEGA's sales Department for a CAL-4 calibration report.
Type K thermocouples have a general temperature range of -200 to 1260°C (-326 to 2300°F), however there are some caveats to this:
Wire Gage | Diameter | Maximum Suggested Temperature |
#8 AWG | 0.128” | 1280°C/2300°F |
#14 AWG | 0.064” | 1090°C/2000°F |
#20 AWG | 0.032” | 980°C/1800°F |
#24 AWG | 0.020” | 870°C/1600°F |
#28 AWG | 0.013” | 870°C/1600°F |
#30 AWG | 0.010” | 760°C/1400°F |
The following is a selection of output values for Type K Thermocouples at selected temperatures. Note: The following output is based on a reference Junction at 0°C/32°F:
Temperature | Output |
0°C | 0.000 mV |
100°C | 4.096 mV |
200°C | 8.138 mV |
300°C | 12.209 mV |
400°C | 16.397 mV |
500°C | 20.644 mV |
600°C | 24.905 mV |
700°C | 29.129 mV |
800°C | 33.275 mV |
900°C | 37.326 mV |
1000°C | 41.276 mV |
1100°C | 45.119 mV |
1200°C | 48.838 mV |
Thermocouples are not manufactured to a specific resistance, they measure temperature based on the output voltage they provide. Resistances will vary with conductor size and somewhat with the lot of conductor used.
The cost of a Type K thermocouple will depend upon the style of the sensor you choose. Once you determine what style sensor you need you can find pricing on our website at omega.com. If you need assistance in selecting the thermocouple you need, please contact our Application Engineering staff who can assist you in your selection.